Personal Monitoring in work places with Dräger Pac 3000, the innovative single gas
instrument, ensures reliable monitoring of ambient air and fast warnings against harmful
concentrations of hydrogen sulfide. Dräger Pac 3000 can be used for over
2 years maintenance-free. Dräger Pac 3000 H2S Measuring Range: 0 to 100 ppm Alarm Threshold A-1: 10 ppm Alarm Threshold A: 20 ppm.
Small and robust;
The small and ergonomic instrument is
covered with a rubber housing and meets
the requirements of IP 65. Size and weight
of Dräger Pac 3000 are perfect for personal
monitoring. An alligator clip provides
secure attachment of the unit to the wearer.
Easy to use:
The two-button-operation is intuitive and
can be locked to avoid misuse. An easy to
change lithium battery powers the instrument.
The IR interface means configuration
and calibrations are easy to perform.
Display;
The large and language-free display indicates
the actual concentration during alarm,
notice and warning functions, and the
remaining operating time. The display has
a strong backlight.
New sensor technology
The new Dräger XXS sensors react rapidly
using very fast electro-chemical reaction
times and a very short diffusion path.
Bump test mode
The instrument is equipped with a bump
test mode. The bump test adjustments are
activated using the PC software, Pac
Dräger Pac 3000:
Handy, reliable with
large display.
Vision or CC Vision. The bump test can be
performed manually or automatically. A
Bump Test Station is available. If the bump
test is required a notice will be indicated in
the display. The bump test period can be
adjusted by the user.
Adjustable operating timer;
With this function an individual operating
period in days can be set, e. g. to adjust a
“calibration due date”, an “inspection due
date”, or a special “out of order date”. The
adjustment can be done by PC with installed
software Pac Vision or CC Vision.
Alarm- / warning functions;
The instrument is equipped with a vibrating
alarm, an acoustic alarm and optical alarm.
For optimal alarm notification, a two-tonealarm
is used. Additionally, before the end of
the instrument’s useful life and before the
end of the battery capacity, a warning will
also occur. Depending on the selected configuration
the pre- or main-alarm can be
acknowledged.
Calibration and configuration
For calibration or individual configuration of
all functions, the Dräger Pac 3000 is connected
to a PC using the connecting cradle
or the E-Cal System.
TECHNICAL DATA
Dimensions (H x W x D) 64 x 84 x 20 (battery compartment: 25) mm (2.5 x 3.3 x 0.8 in., battery compartment: 1 in.)
Weight 106 g (3.8 oz.)
Ambient conditions Temperature -30 to 50 °C (-20 to 120 °F)
Pressure 700 to 1,300 hPa
Humidity 10 to 90 % r. h.
Ingress protection IP 65
Display Language-free LCD display, indication of gas type to be measured, actual concentration
during alarm, operating time, notice and warning functions
Operating time (typical at 25 °C) 2 years
Battery life (typical at 25 °C, 24 hours per CO, H2S: > 10,400 hours
day, 1 minute alarm per day) O2 > 3,600 hours
Acoustic alarm Two-Tone-Alarm, typical > 90 dB (A), at a distance of 30 cm
Approvals CE-Sign (89/336/EEC, 94/9/EC)
ATEX I/II 1/2 G EEx ia I/IIC, T4
UL Class I, Div 1, Group A, B, C, D, E, F, G, Temp. Code T4
cUL Class I, II Div 1, Group A, B, C, D, E, F, G, Temp. Code T4