DrägerSensor® microPac CO P/N 6810030 For microPac, microPac Plus, and X-am 3000
Gases qualified to measure: Carbon Monoxide (CO)
Compatible Instruments: microPac, microPac Plus, and X-am 3000 Monitors
Measuring Range: 0 to 1000 ppm CO
Response Time: 90% response to exposed concentration in less than 20 seconds
Repeatability: +/- 1% of Measured Value
Display Resolution: 1 ppm
Environmental Conditions: Temperature: -40 to 122 °F (-40 to 50 °C), continuous operation
-40 to 140 °F (-40 to 60 °C), short duration
Humidity: 10 to 90 %RH, continuous operation
5 to 99 %RH, short duration
Pressure: 20.7 to 38.4 inches Hg (700 to 1300 hPa)
Note: Rapid changes in environmental conditions may cause temporary fluctuations on the sensor
Cross-Sensitivities: Positive Effects: H2, H2S, SO2, NO, ethanol, and ethylene
Negative Effect: Cl2
Selective Filter: A2T removes H2S and various organic vapors
Note: Other gases may influence the sensor.
Calibration Frequency: Recommended: 12 months
Calibration Gas: Required: Carbon Monoxide
Concentration: 40-100% of full scale range
Minimum: 25 ppm
Compatible Filters: Internal Selective Filter (A2T): P/N 6810378 (provided with sensor)
Internal Dust & Mist Filter: P/N 6809021 (optional)
Warranty Period: 2 Years
The information contained within this publication is Copyright © 1976 - to present, either by the individual manufacturers, The Safety Equipment Store®, or both parties. The Safety Equipment Store®, and, Your One Source For Workplace Safety®, are Registered Service Marks of S&E Industrial Supply Co., Inc. 1976 to present.