DrägerSensor® microPac O2 P/N 6810034
Gases qualified to measure: Oxygen (O2)
Compatible Instruments: microPac Plus and X-am 3000 Monitors
Measuring Range: 0 to 25.0 % Vol. O2
Response Time: 90% response to exposed concentration in less than 20 seconds
Repeatability: +/- 1% of Measured Value or better
Display Resolution: 0.1 %Vol.
Environmental Conditions: Temperature: -40 to 122 °F (-40 to 50 °C), continuous operation
-40 to 140 °F (-40 to 60 °C), short duration
Humidity: 10 to 95 %RH, continuous operation
5 to 99 %RH, short duration
Pressure: 20.7 to 38.4 inches Hg (700 to 1300 mbar)
Note: Rapid changes in environmental conditions may cause temporary fluctuations on the sensor
Cross-Sensitivities: No significant interference by other gases in the TLV range. Larger quantities of other gases may displace the oxygen in the ambient atmosphere causing the sensor readings to drop.
Calibration Frequency: Recommended: 12 months
Calibration Gas: Recommended: Oxygen or Zero Air
Fresh-Air Adjustment (when applicable) is acceptable for most applications
Concentration: 40 to 100 % of measuring range
Minimum: 5.0 % Vol.
Compatible Filters: Internal Dust & Mist Filter: P/N 6809021 (provided with sensor)
Warranty Period: 2 Years
|